Simultaneously detects both total coliforms and Escherichia coli in water, or fecal coliforms in wastewater, giving you results in 18 hours.
Minimises evening and weekend work.
Can be used for presence/absence (P/A) or quantification testing with Quanti-Tray and Quanti-Tray/2000.
The Colilert-18 test uses a proprietary Defined Substrate Technology (DST) nutrient indicators ONPG and MUG to detect coliforms and E. coli. Coliforms use their β-galactosidase enzyme to metabolize ONPG and change it from colorless to yellow.
E. coli use β-glucuronidase to metabolize MUG and create fluorescence. Since most non-coliforms do not have these enzymes, they are unable to grow and interfere. The few non-coliforms that do have these enzymes are selectively suppressed by the Colilert-18 test’s specifically formulated matrix.
This approach is different from traditional media, which provide a nutrient-rich environment that supports the growth of both target organisms and nontargets. When nontargets grow and mimic target organisms, false positives occur. Growth of nontargets can also suppress target organisms and give false negatives in traditional media. To suppress nontargets, traditional media often include high levels of salts, detergents, or other selective agents that may inadvertently suppress target organisms and give further false negatives.
Only IDEXX reagents have been validated and are approved for use with Quanti-Tray and Quanti-Tray/2000.
Learn how to use the Colilert test
Add reagent to sample and incubate at 35°C ± 0.5°C for 18 hours.
Read results:
Colorless = negative
Yellow = total coliforms
Yellow/fluorescent = E. coli
Add reagent to sample.
Pour into Quanti-Tray (counts from 1–200) or Quanti-Tray/2000 (counts from 1–2,419)
Seal in Quanti-Tray Sealer and place in 35°C ± 0.5°C incubator for 18 hours (in other countries, the temperature requirement may be different per regulatory requirements).
Quanti-Tray—Read results:
Quanti-Tray/2000—Read results:
Please reach us at vega@aguasegura-ao.com if you cannot find an answer to your question.
Yes. The International Organization for Standardization (ISO) has withdrawn ISO 9308-2:1990, which follows the multiple-tube (most probable number) method, and replaced it with Colilert-18/Quanti-Tray, the new, worldwide ISO standard 9308-2:2012 for detecting total coliforms and E. coli in water.
Yes. The Colilert-18 test is U.S. Environmental Protection Agency (EPA) approved for several types of regulated water including:
The Colilert-18 test detects coliforms and E. coli at 1 organism/100 mL.
The Colilert-18 test shelf life is up to 15 months from date of manufacture.
The powder should be off-white in color, dry, and free flowing. A slight tinge of color may be observed when the Colilert-18 test is added to the water sample, but this in no way interferes with the “readability” of the test. Any concerns about the color or integrity of the powder should be directed to IDEXX Technical Service.
Yes, presence/absence (P/A) samples need to be prewarmed; however, Quanti-Tray or Quanti-Tray/2000 samples do not. For P/A, if the sample is not already at 33°C–38°C, place the sample in a 35°C waterbath for 20 minutes or a 44.5°C waterbath for 7–10 minutes. This prewarming time is part of (not in addition to) the 18-hour incubation period.
Yes, if an inoculated Colilert-18 test sample is inadvertently incubated beyond 18 hours, the result is stable for 4 additional hours (22 hours maximum incubation). Heterotrophic organisms are suppressed during the incubation. After 22 hours, a negative test is valid, but any yellow color greater than or equal to the comparator is not valid and should be repeated or verified.
Yes, the Colilert-18 test can be run with Quanti-Tray, Quanti-Tray/2000, or any multiple-tube format.
Yes, please refer to the Colilert-18 test product insert for the fecal coliform procedure.
Check with your local, state, and/or federal authorities for proper disposal of bacteriological biohazard materials at your facility.
2°C–25°C away from light.